DHS 157.78(10)(a)4.
4. The operator shall ensure there are no bystanders within a radius of at least 2 meters (6.5 ft) from the patient being examined during exposures.
DHS 157.78(10)(b)1.
1. Hand-held intraoral dental radiographic units shall be used for patient examinations in dental offices that meet the structural shielding requirements specified by the department.
DHS 157.78(10)(b)2.
2. Hand-held intraoral dental radiographic units may not be used for patient examinations in hallways and waiting rooms.
DHS 157.78 History
History: CR 01-108: cr.
Register July 2002 No. 559, eff. 8-1-02;
CR 16-078: am. (4) (d) 1., 3., (8) (title), (9) (a), cr. (10)
Register January 2018 No. 745, eff. 2-1-18.
DHS 157.79
DHS 157.79 Veterinary medicine x-ray systems. DHS 157.79(1)(1)
General. The requirements of this section apply to all animal use x-ray systems used in veterinary practice and are in addition to other provisions in subchs.
I and
III.
DHS 157.79(2)(a)(a) The tube housing shall be electrically shock proof and of a diagnostic type. The x-ray tube may not be hand-held during exposures.
DHS 157.79(2)(b)
(b) A device shall be provided to terminate the exposure after a preset time or exposure.
DHS 157.79(2)(c)
(c) A dead-man type of exposure switch shall be provided with an electrical cord of sufficient length so that the operator or the assistant, may stand out of the useful beam and at least 2 meters (6.5 feet) from the table during all x-ray exposures. A foot operated exposure switch may be used and this switch may be integrated into the table base or the foot switch may be on a 2 meter (6.5 feet) cord.
DHS 157.79(3)(a)(a) The operator shall stand at least 2 meters (6.5 feet) from the tube housing and the animal during radiographic exposures. The operator may not stand in the useful beam. Hand-held fluoroscopic screens may not be used. The tube housing may not be held by the operator. No person other than the operator may be in an x-ray room while exposures are being made unless another person's assistance is required.
DHS 157.79(3)(b)
(b) During any application in which the operator is not located behind a protective barrier, the operator and any other persons in the room during exposures shall wear protective clothing consisting of a protective apron having a lead equivalent of not less than 0.5 millimeter unless measurements indicate otherwise.
DHS 157.79(3)(c)
(c) Any person holding or supporting an animal or the image receptor during radiation exposure shall wear protective gloves that surround the hand and a protective apron having a lead equivalent of not less than 0.25 millimeter. Devices that only partially shield the hands are prohibited.
DHS 157.79(3)(d)
(d) Veterinary fluoroscopy systems shall be operated only under the direct supervision of the licensed veterinarian.
DHS 157.79(4)
(4)
Animal support. Mechanical restraints shall be used to restrict movement of the animal unless the restraints interfere with the examination of the animal. No persons may be regularly utilized to hold or support animals during radiation exposures. Operating personnel may not perform this service except in cases where no other person is available.
DHS 157.80
DHS 157.80 Computed tomography x-ray systems. DHS 157.80(1)(1)
Equipment requirements. A computed tomography (CT) x-ray system shall meet all of the following requirements, as applicable:
DHS 157.80(1)(a)1.1. Means shall be provided to terminate the x-ray exposure automatically by either de-energizing the x-ray source or shuttering the x-ray beam in the event of equipment failure affecting data collection. The termination shall occur within an interval that limits the total scan time to no more than 110% of its preset value through the use of either a back-up timer or devices which monitor equipment function. A visible signal shall indicate when the x-ray exposure has been terminated. The operator shall be able to terminate the x-ray exposure at any time during a scan, or series of scans under CT x-ray system control, of greater than one-half second duration.
DHS 157.80(1)(a)2.
2. For systems that allow high voltage to be applied to the x-ray tube continuously and that control the emission of x-ray with a shutter, the radiation emitted may not exceed 0.88 mGy (100mR) in 1 hour at any point 5 centimeters outside the external surface of the housing of the scanning mechanism when the shutter is closed.
DHS 157.80(1)(a)3.
3. Compliance shall be determined by measurements averaged over an area of 100 square centimeters with no linear dimension greater than 20 centimeters.
DHS 157.80(1)(b)1.
1. A single tomogram system shall allow for visual determination of the tomographic plane or a reference plane offset from the tomographic plane.
DHS 157.80(1)(b)2.
2. A multiple tomogram system shall allow for visual determination of the location of a reference plane.
DHS 157.80 Note
Note: The reference plane may be offset from the location of the tomographic planes.
DHS 157.80(1)(b)3.
3. If a device using a light source is used to satisfy the requirements in subd.
2., the light source shall provide illumination levels sufficient to permit visual determination of the location of the tomographic plane or reference plane under ambient light conditions of up to 500 lux.
DHS 157.80(1)(c)
(c)
Beam-on and shutter status indicators and control switches. DHS 157.80(1)(c)1.1. The CT x-ray control and gantry shall provide visual indication whenever x-rays are produced and, if applicable, whether the shutter is open or closed. Each emergency button or switch shall be clearly labeled as to its function.
DHS 157.80(1)(c)2.
2. For systems that allow high voltage to be applied to the x-ray tube continuously and that control the emission of x-ray with a shutter, the radiation emitted may not exceed 0.88 mGy (100 mR) in one hour at any point 5 cm outside the external surface of the housing of the scanning mechanism when the shutter is closed.
DHS 157.80(1)(c)3.
3. Compliance shall be determined by measurements averaged over an area of 100 square cm with no linear dimension greater than 20 cm.
DHS 157.80(1)(d)
(d)
Indication of CT conditions of operation. A CT x-ray system shall be designed such that the CT conditions of operation to be used during a scan or a scan sequence shall be indicated prior to the initiation of the scan or a scan sequence. On equipment having all or some of these conditions of operation at fixed values, this requirement may be met by permanent markings. Indication of CT conditions of operation shall be visible from any position from which scan initiation is possible.
DHS 157.80(1)(e)
(e)
Maximum surface CTDI100 identification. The angular position where the maximum surface CTDI100 occurs shall be identified to allow for reproducible positioning of a CT dosimetry phantom.
DHS 157.80(1)(f)
(f)
CT x-ray systems containing a gantry manufactured after September 3, 1985. A computed tomography x-ray system containing a gantry that was manufactured after September 3, 1985, shall meet all the following requirements:
DHS 157.80(1)(f)1.
1. The total error in the indicated location of the tomographic plane or reference plane may not exceed 5 millimeters.
DHS 157.80(1)(f)2.
2. If the x-ray production period is less than one-half second, the indication of x-ray production shall be actuated for at least one-half second. Indicators at or near the gantry shall be discernible from any point external to the patient opening where insertion of any part of the human body into the primary beam is possible.
DHS 157.80(1)(f)3.
3. The deviation of indicated scan increment versus actual increment may not exceed plus or minus one millimeter with any mass from 0 to 100 kilograms resting on the support device. The patient support device shall be incremented from a typical starting position to the maximum incremented distance or 30 centimeters, whichever is less, and then returned to the starting position. Measurement of actual versus indicated scan increment may be taken anywhere along this patient support device movement distance.
DHS 157.80(1)(f)4.
4. Premature termination of the x-ray exposure by the operator shall necessitate resetting of the CT conditions of operation prior to the initiation of another scan.
DHS 157.80(1)(f)5.
5. Two-way aural communication shall exist between the patient and the operator at the control panel.
DHS 157.80(1)(f)6.
6. A viewing window, or viewing system, such as closed circuit television or an equivalent, shall be installed to permit continuous observation of the patient during irradiation. The window or viewing system shall be installed such that the operator, located at the control position, can continuously observe the patient during irradiation. When the primary viewing system is electronic, an alternative system shall be available for use in the event of failure of the primary viewing system.
DHS 157.80(2)(a)1.1. A CT x-ray system for human use may only be operated for diagnostic procedures by an American registry of radiologic technologists certified person who is licensed as a radiographer by the State of Wisconsin or has met the radiographer license exemptions and has been specifically trained in its operation.
DHS 157.80(2)(a)2.
2. A CT x-ray system for veterinary use may only be operated for diagnostic procedures by a person who is certified by the American registry of radiological technologists or has completed training equivalent to the requirements of ch.
DHS 157 Appendix L and has been specifically trained in its operation.
DHS 157.80(2)(a)3.
3. Combination systems which combine CT with another imaging device that uses radioactive material shall be operated by a person qualified by training in the safe use of radioactive materials and who meets the training requirements of ch.
DHS 157 Appendix L.
DHS 157.80(2)(b)
(b) Information shall be available at the control panel regarding the operation and calibration of the system. The information shall include all of the following components:
DHS 157.80(2)(b)1.
1. Dates of the latest calibration and spot checks and the location within the facility where the results of those tests may be obtained.
DHS 157.80(2)(b)2.
2. Instructions on the use of the CT dosimetry phantom including a schedule of spot checks appropriate for the system, allowable variations for the indicated parameters and the results of at least the most recent spot checks conducted on the system.
DHS 157.80(2)(b)3.
3. The distance in millimeters between the tomographic plane and the reference plane if a reference plane is utilized.
DHS 157.80(2)(b)4.
4. A current technique chart or list of protocols available at the control panel, which specifies for each routine examination the CT conditions of operation and the number of scans per examination including body part size and correct kV/mA for that body part. The technique chart or a list of protocols shall be used to adjust techniques based on the body part being examined.
DHS 157.80(2)(c)
(c) Calibration and spot check measurements shall be made at a frequency recommended by the manufacturer or established by a medical physicist. If the calibration or spot check of the CT x-ray system identifies that a system operating parameter has exceeded a tolerance established by the medical physicist, use of the CT x-ray system on patients shall be limited to those uses permitted by established written instructions of the medical physicist.
DHS 157.80(2)(d)
(d) A facility shall follow the manufacturer's daily start up routines and preventative maintenance schedules for a specific computed tomography x-ray system.
DHS 157.80 History
History: CR 01-108: cr.
Register July 2002 No. 559, eff. 8-1-02;
CR 06-021: renum. (1) (a) and (c) to be (1) (a) 1. and (c) 1., cr. (1) (a) 2. and 3. and (c) 2. and 3., am. (2) (a)
Register October 2006 No. 610, eff. 11-1-06;
CR 09-062: renum. (2) (a) to be (2) (a) 1. and 3. and am., cr. (2) (a) 2.
Register April 2010 No. 652, eff. 5-1-10;
CR 16-078: r. (1) (b) (intro.), cr. (1) (f) 5., 6., am. (2) (a) 1., (b) 4., (c)
Register January 2018 No. 745, eff. 2-1-18; correction in (2) (a) 2., 3. made under s.
35.17, Stats.,
Register January 2018 No. 745;
CR 22-015: am. (1) (c) 2. Register June 2023 No. 810, eff. 7-1-23.
DHS 157.81(1)(1)
Plan review and approval. Prior to construction, the floor plans, shielding specifications and equipment arrangement of all new installations, or modifications of existing installations, utilizing ionizing radiation machines, including dental CT and dental cephalometric machines, shall be submitted to the department for review and approval.
DHS 157.81 Note
Note: Plans may be mailed to the department at: Department of Health Services, Radiation Protection Section, P.O. Box 2659, Madison WI 53701-2659 or package delivery to: Department of Health Services, Radiation Protection Section, Room 150, 1 West Wilson St, Madison WI 53702-0007.
DHS 157.81(2)
(2)
Exemptions. Dental intraoral and panoramic, mammography, and bone density devices are exempt from this section.
DHS 157.81(3)(a)(a) A shielding plan for a facility with two or more x-ray rooms shall include a medical physicist or person approved by a medical physicist recommendation for shielding.
DHS 157.81(3)(b)
(b) A shielding plan submitted for department review shall include all of the following:
DHS 157.81(3)(b)2.
2. A scale drawing of the room in which a stationary radiation machine system is located with such drawing indicating the use of areas adjacent to the room and an estimation of the extent of occupancy by a person in such areas. In addition, the drawing shall include all of the following:
DHS 157.81(3)(b)2.a.
a. The type and thickness of materials, or lead equivalency, of each protective barrier.
DHS 157.81(3)(b)2.b.
b. The use and occupancy of the areas surrounding an x-ray room, including occupied areas above or below an x-ray room.
DHS 157.81(3)(c)
(c) The operator booth described in the shielding plan shall be designed to meet all the following requirements:
DHS 157.81(3)(c)1.
1. The view area of the window shall be at least 0.09 m
2 (144 square inches).
DHS 157.81(3)(c)2.
2. The window shall be placed so that the edge of the view window is at least 0.45 meters (18 inches) from the end of the barrier. The window shall be placed so that the patient may be observed at all times and each entrance to the room is observed from the operator position. Patient and entrance observation may be accomplished by the use of electronic devices or mirrors.
DHS 157.81(3)(c)3.
3. The shielding value of the window shall be equal to the wall in which it is mounted.
DHS 157.81(3)(c)4.
4. Booth walls shall be 2.1 meters (7 feet) in height and permanently attached to the floor or walls. The booth shall be at least 1.3 meters (4 feet) from the nearest vertical cassette holder or 0.3 meters (one foot) from the nearest corner of the examining table.
DHS 157.81(3)(c)5.
5. When a door or moveable panel is used as an integral part of the booth structure, it shall have a permissive device that prevents an exposure when the door or panel is not closed.
DHS 157.81(3)(c)6.
6. Verbal communication with the patient shall be possible at all times during the x-ray procedure.
DHS 157.81(3)(c)7.
7. The x-ray exposure control shall be located within the shielded area and at least 1 meter (3.3 feet) from the open end of the protective barrier, excluding mammography units.
DHS 157.81(4)
(4)
Operational analysis. The department may require additional modifications to a shielding plan after initial approval of the plan if a subsequent analysis of operating conditions indicates the possibility of a person receiving a dose in excess of the limits prescribed in ss.
DHS 157.22 (1) and
(5) to
(8) and
157.23 (1) and
(2). An existing x-ray room constructed using 5 mSv (500 mrem) as the public exposure limit may continue to operate without modification until the x-ray equipment is replaced or the room is modified.
DHS 157.81 History
History: CR 01-108: cr.
Register July 2002 No. 559, eff. 8-1-02;
CR 06-021: am. (3) (a)
Register October 2006 No. 610, eff. 11-1-06;
CR 09-062: am. (1) and (2), cr. (3) (c) 7.
Register April 2010 No. 652, eff. 5-1-10;
CR 22-015: am. (4) Register June 2023 No. 810, eff. 7-1-23. DHS 157.82
DHS 157.82 General administrative requirements for facilities using therapeutic radiation machines for human use. DHS 157.82(1)(1)
Administrative controls. A registrant shall be responsible for directing the operation of the therapeutic radiation machines that have been registered with the department. All persons operating a therapeutic radiation machine shall be instructed in and shall comply with the provisions of the radiation safety requirements of ss.
DHS 157.22 (1) and
(5) to
(8) and
157.25 (2). A therapeutic radiation machine that does not meet the provisions of this subchapter but is of a type accepted by the US food and drug administration may not be used for irradiation of human patients.
DHS 157.82(2)(a)
(a) A registrant for any therapeutic radiation machine, except dermatology units under 150 kV, shall require the authorized user to be a physician who meets any of the following requirements:
DHS 157.82(2)(a)1.c.
c. Radiology, with specialization in radiotherapy, as a British “Fellow of the Faculty of Radiology" or “Fellow of the Royal College of Radiology".
DHS 157.82(2)(a)1.d.
d. Therapeutic radiology by the Canadian royal college of physicians and surgeons.
DHS 157.82(2)(a)2.
2. Actively practices therapeutic radiology and has completed all of the following:
DHS 157.82(2)(a)2.b.
b. Two hundred hours of instruction in basic radiation techniques applicable to the use of an external beam radiation therapy unit.